Pacemaker Rhythms and Artifact Recognition
Objective. Students will identify pacemaker spikes, determine pacemaker capture, and distinguish common artifacts (somatic tremor, AC interference, lead wander, loose electrode) from true dysrhythmias on rhythm strips.
⚠ Pending professional review
This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.
VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.
What Is a Pacemaker?
A cardiac pacemaker is an implanted device that delivers a small electrical impulse to the heart when the heart's own conduction system fails to do so reliably. Pacemakers are used when patients have symptomatic bradycardia, heart block, or other conditions in which the natural rate or rhythm is dangerously slow or irregular. As an EKG technician, you will frequently record rhythms from patients who have pacemakers. You must recognize pacemaker activity, confirm that it is working correctly, and know when something looks wrong — so the appropriate clinician can be notified promptly.
The Pacemaker Spike
Every pacemaker stimulus produces a sharp, narrow vertical mark on the rhythm strip called a pacemaker spike (also called a pacing artifact). This spike is the electrical footprint of the device firing. Spikes are thin and vertical and stand out clearly from the surrounding waveforms. Your first task on any strip is to locate these spikes.
- Atrial pacing: The spike appears just before a P wave. The QRS that follows should be narrow and normal-looking because ventricular conduction continues through the natural pathway.
- Ventricular pacing: The spike appears just before a QRS complex. Because the impulse bypasses the normal conduction pathway, the resulting QRS is wide and has an abnormal, bundle-branch-block-like appearance. There may be no visible P wave before the spike, or the P wave is unrelated to the paced beat.
- Dual-chamber (AV) pacing: Two spikes appear per cycle — one before the P wave (atrial spike) and one before the QRS (ventricular spike). This mimics the heart's natural sequential contraction of atria then ventricles.
Determining Capture
Capture means the pacemaker's stimulus actually caused the heart muscle to depolarize. Look for a waveform immediately after every spike:
- Atrial capture: A P wave follows the atrial spike.
- Ventricular capture: A wide QRS follows the ventricular spike.
- Failure to capture: A spike appears but NO waveform follows. This is a malfunction that must be reported to the nurse or supervising clinician immediately because the patient may not be getting the pacing support they need.
- Failure to pace (failure to fire): Expected spikes are absent. Also a malfunction requiring immediate escalation.
Note: in a patient who is fully pacemaker-dependent, failure to capture or failure to pace can be life-threatening. Never delay reporting these findings.
Understanding Sensing
Most pacemakers are programmed to sense the patient's own heartbeats. When the device detects a native beat, it resets its timer and holds off firing — this is called inhibition. You may see a mix of native beats and paced beats on the same strip. A paced beat that fires when it should be inhibited (competes with a native beat) is called pacemaker competition or failure to sense, and it should also be reported.
Common EKG Artifacts and How to Recognize Them
Artifacts are signals on the strip that come from sources other than the heart. They can mimic dysrhythmias or obscure real rhythms. Recognizing them protects patients from misdiagnosis and unnecessary treatment.
Somatic Tremor (Muscle Tremor)
Caused by patient movement, shivering, Parkinson's disease, or anxiety. Produces an irregular, chaotic, high-frequency baseline that can look like atrial fibrillation or even ventricular fibrillation. Key clue: you can still find regular R waves marching through the noise if you look carefully.
AC (Alternating Current) Interference
Caused by nearby electrical equipment, poor grounding, or a lead wire touching a metal bed rail. Produces a perfectly regular, fine-tooth-comb waveform at 60 cycles per second (60 Hz in the US). The baseline thickens into a uniform fuzzy band. Correction: move or turn off nearby electrical equipment; check lead wire routing; ensure the machine is grounded.
Baseline Wander (Baseline Drift)
The baseline rolls up and down in a slow wave, often caused by patient breathing, patient movement, or electrode gel drying out. Correction: ask the patient to breathe quietly and avoid talking; replace dried electrodes; ensure lead wires are not pulling on electrodes.
Loose or Detached Electrode (Lead-Off Artifact)
Produces a straight flat line or wildly erratic signal in one lead. Other leads on the same strip often look normal — a key clue that the problem is mechanical, not cardiac. Correction: check and reattach the electrode; clean and dry the skin; replace the electrode if needed.
Safety and Scope of Practice
- Never interpret a rhythm strip as a diagnosis or tell a patient they are having a heart attack or that their pacemaker is broken. Your role is to acquire and document the tracing and report abnormalities to the supervising clinician.
- Always report suspected failure to capture or failure to pace immediately — do not wait until end of shift.
- If a strip shows what could be ventricular fibrillation, confirm all leads are attached before concluding it is artifact. Patient safety depends on ruling out true life-threatening rhythms first.
- Wash hands and follow facility infection-control policy when placing and removing electrodes.
- Document what you observe on the strip (e.g., "pacemaker spikes present, capture noted") and any corrective actions you took for artifacts.
Key terms9
Pacemaker spikeA sharp, narrow vertical deflection on the EKG strip produced each time the pacemaker fires an electrical stimulus.
CaptureSuccessful depolarization of the heart muscle following a pacemaker spike, evidenced by a P wave (atrial) or a wide QRS (ventricular) immediately after the spike.
Failure to captureA pacemaker malfunction in which a spike is visible on the strip but no waveform follows, indicating the stimulus did not depolarize the myocardium.
Dual-chamber pacingA pacing mode in which the device paces both the atrium and the ventricle sequentially, producing two spikes per cardiac cycle — one before the P wave and one before the QRS.
Somatic tremorArtifact caused by skeletal muscle activity (movement, shivering, tremor) that produces an irregular, high-frequency, chaotic baseline that can resemble certain dysrhythmias.
AC interferenceA regular 60 Hz artifact from nearby electrical sources that creates a uniform, fine, thickened baseline resembling a comb; corrected by removing electrical interference or improving grounding.
Baseline wanderA slow, rolling movement of the EKG baseline caused by patient breathing, movement, or poor electrode contact, which can obscure ST-segment changes and other features.
Failure to senseA pacemaker malfunction in which the device does not detect the patient's native heartbeats and fires inappropriately, potentially competing with intrinsic beats.
ArtifactAny signal on the EKG strip that does not originate from the heart's electrical activity; artifacts can mimic or obscure true cardiac rhythms.
In class: Instructor demonstrates pacemaker spike identification on a rhythm monitor trainer, showing atrial, ventricular, and dual-chamber pacing with and without capture; students annotate spikes and captured beats on printed strips. · Artifact gallery walk: eight stations around the room each display an enlarged strip showing one artifact type; students rotate, identify the artifact at each station on a recording sheet, and list one corrective action. · Paired troubleshooting exercise: one student operates an EKG practice trainer to introduce deliberate artifact (by moving the lead wire); partner identifies artifact type and states corrective steps; roles switch.
Materials: Rhythm monitor trainer or EKG simulator, printed pacemaker strip set, artifact gallery station sheets, recording worksheets, EKG practice trainer or simulator with lead wires
Check: Ten-item written and strip-based quiz: five pacemaker rhythm strips (identify type and capture status) and five artifact strips (name artifact and correction); 80% required to pass.
Practice check8