newSchoolWorks.com › VoTechWorks › Health & Personal Services › Limited X-Ray / Radiologic Technology › Radiographic Image Evaluation › Lesson 6
HLT-RADI64-04 · Lesson 6 · 26 clock hours · theory

Radiographic Artifacts: Recognition and Source Identification

Objective. Students classify radiographic artifacts by source category—patient-related, equipment-related, detector-related, and processing-related—and distinguish artifacts from anatomical structures or pathological findings.

⚠ Pending professional review

This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.

VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.

What Is a Radiographic Artifact?

A radiographic artifact is any density or structure that appears on an image but does not correspond to a real anatomical structure in the patient. Artifacts can mimic disease, obscure pathology, or simply reduce diagnostic quality. Because an unrecognized artifact can lead to misdiagnosis or a repeated exposure, identifying and eliminating artifacts is a core radiographer skill and a direct patient-safety responsibility.

Artifacts fall into four major source categories: patient-related, equipment-related, detector-related, and processing-related. Learning to classify an artifact by its source tells you both why it occurred and how to prevent it.

---

Patient-Related Artifacts

Patient-related artifacts originate with the patient's body or actions before or during the exposure.

Safety point: Before every exposure, perform a thorough patient assessment and ask about implants, piercings, and recent contrast studies. Never assume the patient knows what must be removed.

---

Equipment-Related Artifacts

Equipment artifacts arise from the X-ray tube, generator, Bucky, or grid.

---

Detector-Related Artifacts

In computed radiography (CR) and digital radiography (DR), the image receptor itself can be the source.

---

Processing-Related Artifacts

In facilities that still use film-screen or wet-processor systems, processing artifacts occur.

In digital environments, PACS display settings (incorrect window/level, monitor artifacts) can create pseudo-artifacts. Always evaluate images on a calibrated diagnostic monitor or PACS workstation.

---

Distinguishing Artifacts from Pathology

The key questions to ask:

1. Does the finding follow anatomical boundaries? True pathology usually does. A perfectly straight line crossing bone and soft tissue is almost certainly an artifact.

2. Does the finding repeat position-to-position? A dead detector element appears in the same pixel location on every image from that detector.

3. Is the finding outside the patient's body outline? Artifacts from external objects may project partially outside the collimated field.

4. Can the finding be explained by a known artifact mechanism? If yes, check the source and repeat if clinically necessary.

Safety point: When in doubt, consult the supervising radiologist or radiographer before repeating. Do not repeat exposures on your own authority without following your facility's repeat/reject protocol.

---

Key Takeaway

Artifact recognition protects patients from unnecessary repeat exposures, prevents misdiagnosis, and ensures image quality. Classify first (source category), then identify the mechanism, then correct the cause. Document all repeated exposures in the facility's reject analysis log.

Key terms8

ArtifactAny density or structure visible on a radiographic image that does not correspond to actual patient anatomy; can mimic or obscure pathological findings.
Grid cutoffUnderexposure or darkening of part of the image caused by misalignment of the X-ray beam to the grid's focal distance or angle, preventing grid interspaces from transmitting primary radiation efficiently.
Ghosting (residual image)A faint remnant of a previous image superimposed on a new CR image because the imaging plate was not completely erased before reuse.
Pi linesThin lines running perpendicular to film travel through the processor, spaced at intervals equal to the circumference of a processor roller, caused by a dirty or damaged roller.
Phosphor plate fogIncreased background density on a CR image caused by unintended exposure of the imaging plate to scatter or background radiation before the intended exposure.
Dead detector elementA non-functioning pixel or group of pixels in a digital radiography flat-panel detector that consistently appear as bright or dark spots in the same location on every image.
Motion artifactBlurring or streaking of anatomical structures on an image caused by patient movement, voluntary or involuntary, during the X-ray exposure.
Reject/repeat analysisA quality management process in which repeated or rejected radiographic exposures are documented, categorized by cause (including artifacts), and analyzed to identify patterns and reduce patient dose.
In class: Illustrated lecture presenting at least 15 artifact types with representative images; students complete a classification table sorting each artifact into its source category and noting the visual appearance. · Unknown artifact exercise: instructor presents 10 cropped artifact images without labels; students write the artifact name, likely cause, and corrective action for each. · Lab artifact hunt: students introduce deliberate artifacts (lead object on the cassette, motion via phantom movement, grid cutoff by misaligning the Bucky) and photograph or save the resulting images for their lab log with written analysis.
Materials: PACS workstation, curated artifact image library, CR cassette, grid, lead markers, radiographic phantom, lab log template, Papp Quality Management in the Imaging Sciences
Check: Artifact classification table and unknown artifact exercise; pass requires ≥75% correct on the unknown artifact exercise (≥8 of 10 correct name and cause).

Practice check8

Answer each question, then check it. Nothing is recorded; this is your own practice.

Image Production and Evaluation
1. A radiographer notices a series of thin, evenly spaced lines running across a film at regular intervals. The spacing matches the circumference of the processor rollers. This artifact is most accurately called:
Patient Care and Education
2. A patient's chest radiograph shows a sharply defined, dense oval opacity overlying the right lung base. Before concluding this is a pulmonary mass, the radiographer should first:
Image Production and Evaluation
3. During a digital radiography (DR) exam, the same cluster of bright dots appears in the same location on every image regardless of patient positioning. The most likely cause is:
Image Production and Evaluation
4. A CR image shows a faint superimposed image of a previous patient's hand over the current image of the chest. What is the most likely cause?
Image Production and Evaluation
5. Grid cutoff is most likely to occur when:
Image Production and Evaluation
6. A radiography student accidentally stores a loaded CR cassette in the hallway adjacent to an active X-ray room for several hours before use. The resulting image most likely shows:
Patient Care and Education
7. Which of the following is the BEST first step when a radiographer suspects an artifact is mimicking a pathological finding?
Safety
8. In a reject/repeat analysis program, why is it important to categorize repeated exposures by artifact type?