Objective. Students will perform flat-field uniformity, artifact evaluation, and exposure indicator (EI) calibration checks on digital radiography (DR) and computed radiography (CR) systems.
This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.
VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.
Digital radiography (DR) flat-panel detectors and computed radiography (CR) imaging plates are complex systems. Unlike film, they do not obviously "look wrong" when performance drifts. A detector that produces images with subtle non-uniformity or an exposure indicator (EI) that is drifting out of range can lead to repeated exposures, missed diagnoses, or unnecessary patient dose — all of which affect patient safety and your professional standing. Routine quality control (QC) is not optional; it is a regulatory and accreditation expectation, and understanding it is tested on the ARRT examinations.
Flat-field uniformity testing evaluates whether every pixel across the detector responds equally to a uniform beam of radiation. To perform this test, the detector is exposed without a phantom in the beam — the primary beam alone strikes the entire active area. The resulting image should appear uniformly gray from edge to center.
After acquisition, the image is displayed on a diagnostic-quality monitor and image-processing software is used to sample pixel values across the detector surface. A region-of-interest (ROI) tool is placed in multiple locations — typically the four quadrants and the center. Pixel values (often expressed as signal-to-noise ratio, mean pixel value, or a manufacturer-specific unit) are compared. Significant variation between regions indicates a uniformity problem.
Dead pixels or clusters of dead pixels appear as bright or dark spots that do not respond to exposure. Most manufacturers allow a small number of isolated dead pixels; clusters or lines of dead pixels are cause for service. All findings — acceptable or not — are recorded in the QC log with date, technique used, and technologist initials.
Safety note: Keep your personal dosimeter on and outside the primary beam. Flat-field exposures use live radiation; follow all standard radiation protection practices.
Artifacts are any unwanted structures or patterns visible in the image that are not part of the patient's anatomy. In digital imaging, common artifact sources include:
To perform a structured artifact survey, expose a uniform PMMA (acrylic) phantom under standard technique. The phantom provides a featureless, uniform signal. Review the image systematically by dividing it into quadrants. Record every artifact: its type, its location (using quadrant notation or coordinates), and the most likely cause. A finding is documented even if it is within acceptable limits, because trending over time catches developing problems before they affect clinical images.
The exposure indicator (EI) is a numerical value generated by the imaging system to reflect the amount of radiation that reached the detector. Different manufacturers use different scales and names (e.g., Exposure Index, S-number, lgM), but the International Electrotechnical Commission (IEC) established a standardized EI system that many modern systems follow. Under the IEC framework, a target EI (EI_T) is established for each examination type, and the deviation index (DI) expresses how far the actual EI is from the target:
DI = 10 × log₁₀(EI_measured / EI_target)
A DI of 0 means the detector received exactly the intended exposure. A DI of +1 means approximately twice the intended exposure reached the detector (overexposure). A DI of −1 means approximately half the intended exposure (underexposure). Most facilities set an acceptable DI range of −1 to +1 for routine work; values outside ±3 are generally considered clinically significant.
For the calibration check, a standard PMMA phantom is exposed at a defined, reproducible technique (set kVp, mAs, SID, and field size). The system's displayed EI or DI is recorded and compared to the facility's established target range. If the value falls outside the acceptable range, the cause must be investigated before the system is used clinically. Common causes include technique errors, detector calibration drift, or automatic exposure control (AEC) malfunction.
Why this matters for patient dose: A consistently elevated DI means patients are receiving more radiation than necessary. A consistently negative DI may mean images are underexposed, tempting repeat exposures. Either pattern requires corrective action.
Every QC test result is entered into the facility's QC log immediately after testing. The log entry includes the date, the specific test performed, the technique used, the measured values, the acceptable range, the pass/fail determination, and the technologist's signature. Instructor or supervisory sign-off is required before the system returns to clinical use if a failure is noted. Accurate, legible documentation is a professional and regulatory obligation — incomplete records are treated the same as a failed test during inspections.
The ARRT Limited Scope and ARRT Radiography (RT-R) examinations both include content on equipment quality control, image quality, and radiation protection. Your state radiologic license requires you to practice within the standards of care set by your facility and state regulations, which include QC compliance. Performing and documenting QC correctly from day one is both a patient-safety habit and a credential-maintenance expectation.
Answer each question, then check it. Nothing is recorded; this is your own practice.
A free tech program — complimentary, as our commitment to get America working. Help us build VoTechWorks. Tell us what works, what doesn’t, and what you want to learn.
Does the school offer financial aid?
No. We don’t offer financial aid because we’re free. No, really: completely free.
Do you offer industry certification?
No. We are not certified yet, so we can’t offer that to students. What we do offer is knowledge and understanding of a trade.
Enroll first in the trade you’re interested in. You get your key, your program opens, and you begin studying.
Enroll free →