Objective. Students will identify the type of artifact present on a tracing, state its most likely cause, and apply the correct corrective technique to produce a clean re-acquisition.
This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.
VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.
An artifact is any mark on an electrocardiogram tracing that does not originate from the heart's electrical activity. Artifacts distort or mimic cardiac waveforms and can lead to misinterpretation—a serious patient safety concern. Your job as a medical assistant is to recognize each type of artifact, find its cause, and correct the problem before handing a tracing to the provider for interpretation. Never submit a tracing you know is artifactual without flagging it.
Somatic tremor appears as a jagged, irregular baseline that varies in amplitude and frequency. It is caused by skeletal muscle activity—shivering, Parkinson's tremor, anxiety, pain, or simple voluntary movement during recording. Because the muscle generates its own small electrical signal, the ECG machine picks it up alongside the heart signal.
Correction:
AC interference appears as a thick, fuzzy, uniform baseline with fine rapid oscillations—typically 60 tiny spikes per second matching the frequency of standard US electrical current. It looks like a wide, blurry line rather than a thin crisp one.
Causes: Poor skin preparation, dried or loose electrodes, electrodes touching or near electrical equipment, patient contact with metal table rails, lead wires lying across power cords, or a malfunctioning machine.
Correction:
Wandering baseline appears as the entire baseline slowly drifting up and down across the paper in a wave-like motion. Unlike somatic tremor, the oscillation is slow and smooth.
Causes: Breathing movement (the chest rises and falls, pulling on electrode cables), poor electrode adhesion, improper electrode placement on areas that move with respiration, or electrode gel drying out.
Correction:
Lead reversal is not a random signal—it is a systematic error in which wires are connected to the wrong electrodes. The most common reversal is the right arm (RA) and left arm (LA) leads.
How it appears: Lead I will show an inverted (negative) deflection where a positive deflection is expected. A classic sign of RA–LA reversal is a negative P wave in lead I combined with an inverted QRS, which can mimic dextrocardia or other serious conditions. Lead reversal can cause a tracing to be misread as a life-threatening arrhythmia or infarction.
Correction:
Most artifacts can be prevented at the preparation step. Follow these steps every time:
1. Select the correct anatomical sites.
2. If the site is hairy, clip—do not shave—excess hair.
3. Clean the site with an alcohol pad.
4. Allow it to dry completely.
5. Apply a fresh, within-expiration-date electrode.
6. Press firmly around the entire adhesive ring.
When you encounter a poor-quality tracing, work through this sequence:
1. Look at the pattern — Is the artifact uniform (AC) or irregular (somatic)? Fast oscillations or slow drift (wandering)? Inverted expected waveforms (reversal)?
2. Identify which leads are affected — All leads equally suggests a systemic cause (AC power, patient movement). One or two leads suggest a local electrode or wire problem.
3. Check the patient first — Is the patient comfortable, warm, and still?
4. Check the equipment second — Electrodes, wires, connections, machine settings.
5. Correct, then re-acquire — Document what you found and what you did.
Following this sequence protects both the patient and the integrity of the clinical record.
Answer each question, then check it. Nothing is recorded; this is your own practice.
A free tech program — complimentary, as our commitment to get America working. Help us build VoTechWorks. Tell us what works, what doesn’t, and what you want to learn.
Does the school offer financial aid?
No. We don’t offer financial aid because we’re free. No, really: completely free.
Do you offer industry certification?
No. We are not certified yet, so we can’t offer that to students. What we do offer is knowledge and understanding of a trade.
Enroll first in the trade you’re interested in. You get your key, your program opens, and you begin studying.
Enroll free →