Oxygen Delivery Systems and Supplemental Oxygen Therapy

Objective. Students will select the correct oxygen delivery device and flow rate for a given patient presentation and correctly assemble, apply, and troubleshoot each device.

⚠ Pending professional review

This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.

VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.

Why Supplemental Oxygen Matters

Oxygen is one of the most frequently used therapeutic agents in emergency medicine. When a patient's tissues do not receive enough oxygen—a condition called hypoxia—cells begin to die within minutes. Your job as an EMT or paramedic is to recognize when supplemental oxygen is indicated, select the right delivery device, set an appropriate flow rate, and apply the device correctly. Using too little oxygen can harm a hypoxic patient; using high-flow oxygen unnecessarily on a patient who does not need it carries its own risks. Match the device to the patient.

Fraction of Inspired Oxygen (FiO2)

Room air contains approximately 21% oxygen. The fraction of inspired oxygen, or FiO2, describes how much of the gas a patient breathes is oxygen. Different delivery devices deliver different FiO2 values depending on flow rate, mask fit, and patient breathing pattern. These values are approximate because actual delivered FiO2 varies with respiratory rate and tidal volume.

Common Oxygen Delivery Devices

Nasal Cannula (NC): A nasal cannula consists of two small prongs that rest inside the nostrils connected to oxygen tubing. It delivers a low-to-moderate FiO2—roughly 24–44% at flow rates of 1–6 liters per minute (L/min). Use it for patients who need mild supplemental oxygen, are breathing adequately, and cannot tolerate a mask (such as patients who feel claustrophobic or are eating). Flow rates above 6 L/min do not meaningfully increase FiO2 and may dry and irritate nasal mucosa.

Simple Face Mask (SFM): The simple face mask covers the nose and mouth and has open exhalation ports on the sides. It delivers approximately 35–60% FiO2 at 6–10 L/min. A minimum flow of 6 L/min is required to flush exhaled carbon dioxide (CO2) from the mask. Never run a simple face mask below 6 L/min.

Partial Rebreather Mask (PRM): A partial rebreather mask adds a small reservoir bag to the simple face mask. The patient rebreathes the first portion of exhaled gas (rich in oxygen) from the reservoir before that gas mixes with room air through the side ports. It delivers approximately 35–60% FiO2 at 6–10 L/min. Keep the reservoir bag at least one-third to one-half inflated at all times.

Non-Rebreather Mask (NRM): The non-rebreather mask is the highest-FiO2 device available without specialized equipment. It adds one-way flutter valves over the exhalation ports and between the mask and reservoir bag. These valves prevent exhaled gas from entering the reservoir. When properly fitted and run at 10–15 L/min, it can deliver FiO2 values approaching 90% or higher in ideal conditions. Keep the reservoir bag fully inflated—never let it fully collapse on inhalation. Use it for any seriously ill or injured patient who may be hypoxic: major trauma, suspected carbon monoxide poisoning, respiratory distress, altered mental status with unknown cause.

Venturi Mask: The Venturi mask uses a color-coded adaptor that entrains a precise, fixed ratio of room air. This produces a consistent, predictable FiO2 regardless of the patient's breathing pattern. Common delivered concentrations range from approximately 24–50%, depending on the adaptor used. The Venturi mask is the preferred device when precise oxygen titration is required, such as in patients with chronic obstructive pulmonary disease (COPD) who may be sensitive to high oxygen levels. Always match the adaptor color to the prescribed FiO2 and set the flow rate printed on that adaptor.

Humidified High-Flow Systems: High-flow nasal cannula and aerosol humidification systems deliver heated, humidified oxygen at higher flow rates and can achieve higher FiO2 values. In the prehospital setting these systems are less common but are encountered in interfacility transport and some advanced EMS systems. Humidification reduces mucosal drying during prolonged oxygen therapy.

Oxygen Cylinders: Safety and Duration

Oxygen is stored as a compressed gas. Cylinders are color-coded green in the United States and must be handled with care—a damaged cylinder can become a projectile. Always secure cylinders upright or in an approved carrier. Keep cylinders away from heat and flames. Never use oxygen near an open flame or in the presence of petroleum-based lubricants.

To open a cylinder, crack the valve briefly away from yourself and others to clear dust, then attach the regulator and open fully. To close, turn the valve clockwise until seated, then bleed remaining pressure from the regulator.

The two cylinder sizes you will use most often in the field are the D cylinder (approximately 350 liters capacity when full) and the E cylinder (approximately 625 liters capacity when full). Full pressure is approximately 2,000 psi for both sizes, though always read the actual gauge.

Estimate remaining cylinder time using this formula:

Duration (minutes) = (Gauge pressure in psi × Cylinder constant) ÷ Flow rate (L/min)

Cylinder constants: D = 0.16, E = 0.28.

Applying Devices Safely

Before applying any device: confirm the cylinder is open, confirm flow before placing the device on the patient, and check that tubing is connected and not kinked. After application, reassess the patient's respiratory rate, effort, and SpO2 (if pulse oximetry is available). A mask that is too loose delivers far less FiO2 than expected. Facial hair, trauma, and anatomy all affect mask seal.

Always have a backup device ready. If a non-rebreather mask fails or is not tolerated, a bag-valve mask with reservoir is the next step for a critically hypoxic patient.

Key terms8

FiO2Fraction of inspired oxygen; the proportion of oxygen in the gas mixture a patient inhales, expressed as a decimal or percentage. Room air FiO2 is approximately 0.21 (21%).
HypoxiaDeficiency of oxygen reaching the body's tissues; a primary indication for supplemental oxygen therapy in emergency care.
Nasal cannulaA low-flow oxygen delivery device with two nasal prongs that delivers approximately 24–44% FiO2 at 1–6 L/min; suitable for patients with mild oxygen needs who are breathing adequately.
Non-rebreather maskA high-flow face mask with a reservoir bag and one-way valves that prevents exhaled gas from re-entering the reservoir, capable of delivering FiO2 approaching 90% or higher at 10–15 L/min.
Venturi maskAn oxygen mask that uses color-coded adaptors to entrain a precise, fixed ratio of room air, delivering a consistent, predictable FiO2; preferred when accurate oxygen titration is required.
Reservoir bagThe collapsible bag attached to partial rebreather and non-rebreather masks that stores oxygen between patient breaths; must remain adequately inflated to function correctly.
Cylinder constantA fixed numeric factor specific to each oxygen cylinder size used in the tank-duration formula; D cylinder = 0.16, E cylinder = 0.28.
Cylinder duration formulaDuration (minutes) = (Gauge pressure in psi × Cylinder constant) ÷ Flow rate (L/min); used to estimate how long an oxygen supply will last at a given flow rate.

Worked examples4

1. An E cylinder reads 1,400 psi on the gauge. You are delivering oxygen at 10 L/min via non-rebreather mask. How many minutes of oxygen remain?
  1. Identify the cylinder constant for an E cylinder: 0.28.
  2. Multiply gauge pressure by the cylinder constant: 1,400 × 0.28 = 392.
  3. Divide by the flow rate: 392 ÷ 10 = 39.2 minutes.
Answer: Approximately 39 minutes of oxygen remain.
2. A D cylinder reads 1,000 psi. You switch to a nasal cannula at 4 L/min. How many minutes of oxygen remain?
  1. Identify the cylinder constant for a D cylinder: 0.16.
  2. Multiply gauge pressure by the cylinder constant: 1,000 × 0.16 = 160.
  3. Divide by the flow rate: 160 ÷ 4 = 40 minutes.
Answer: Approximately 40 minutes of oxygen remain.
3. An E cylinder reads 500 psi. You are running oxygen at 15 L/min via non-rebreather mask. How long until the cylinder is empty?
  1. Identify the cylinder constant for an E cylinder: 0.28.
  2. Multiply gauge pressure by the cylinder constant: 500 × 0.28 = 140.
  3. Divide by the flow rate: 140 ÷ 15 ≈ 9.3 minutes.
Answer: Approximately 9 minutes of oxygen remain — change cylinders immediately.
4. You have a D cylinder at 1,800 psi and are delivering oxygen at 6 L/min via simple face mask. How long will the oxygen last?
  1. Identify the cylinder constant for a D cylinder: 0.16.
  2. Multiply gauge pressure by the cylinder constant: 1,800 × 0.16 = 288.
  3. Divide by the flow rate: 288 ÷ 6 = 48 minutes.
Answer: Approximately 48 minutes of oxygen remain.

The math in this lesson2

Learners compute oxygen time remaining as (1,400 psi x 0.28) / 10 L/min = about 39 minutes.
A-SSE.1N-Q.1
Brush up in the math halls at newSchoolWorks.com:
In class: Device identification station: students move through six stations (nasal cannula, simple face mask, partial rebreather mask, non-rebreather mask, Venturi mask, humidified high-flow system) and identify the device, its approximate delivered FiO2 range, and appropriate clinical indication on a written card. · Assembly and application practice: using manikins and mannequin faces, students assemble each device from components, set the correct flow rate on a practice regulator, and apply the device; a lab partner uses a checklist to evaluate technique. · Cylinder safety and calculations exercise: students practice opening and closing a practice oxygen cylinder, reading the regulator gauge, and calculating estimated remaining cylinder time using the tank pressure and estimated flow rate formula.
Materials: Oxygen delivery devices (full set as listed), practice oxygen regulators, D and E cylinder training props, flowmeters, manikin heads, oxygen delivery competency checklists, calculator
Check: Instructor-evaluated skills checklist for assembly and application of non-rebreather mask and nasal cannula on a manikin; student must complete all critical steps correctly with no critical failures; pass requires zero critical errors.

Practice check8

Answer each question, then check it. Nothing is recorded; this is your own practice.

Airway, Respiration, and Ventilation
1. A patient is conscious, alert, and breathing adequately with mild respiratory distress. He reports feeling claustrophobic when a mask is placed over his face. Which oxygen delivery device is most appropriate?
Airway, Respiration, and Ventilation
2. You are using a non-rebreather mask on a critically ill patient. You notice the reservoir bag is completely collapsing with each breath the patient takes. What is the most appropriate corrective action?
Airway, Respiration, and Ventilation
3. Which oxygen delivery device provides the most precise, consistent FiO2 regardless of the patient's respiratory rate or tidal volume?
Airway, Respiration, and Ventilation
4. An E cylinder currently reads 700 psi. You are delivering oxygen at 10 L/min. Approximately how many minutes of oxygen remain? (E cylinder constant = 0.28)
Airway, Respiration, and Ventilation
5. What is the minimum recommended flow rate for a simple face mask, and why?
Airway, Respiration, and Ventilation
6. A patient with a known history of COPD presents with mild shortness of breath and an SpO2 of 88%. Which device allows you to deliver a controlled, low FiO2 most reliably?
Airway, Respiration, and Ventilation
7. When opening a compressed oxygen cylinder, what is the correct first step before attaching the regulator?
Airway, Respiration, and Ventilation
8. A patient is found unresponsive after a house fire. Which oxygen delivery device is most appropriate for immediate use?
Record this lesson on my transcript. Sign in with your key and enrollment email to record lessons (enroll free if you have no key). This page stays open to read and practise.