AGR-WATE71-02 · Lesson 3 · 10.5 clock hours · lab
Turbidity, Suspended Solids, and Color
Objective. Students will measure turbidity in NTU using a nephelometer and total suspended solids by gravimetric method, and relate results to treatment performance standards.
⚠ Pending professional review
This lesson, its worked examples and its practice questions has not yet been verified by a licensed professional in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it as instructions for work on a job site or on live equipment. Always follow your instructor, your supervising professional, and current official references and regulations.
VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.
Why Turbidity, Suspended Solids, and Color Matter
Water leaving a treatment plant must be clear, safe, and aesthetically acceptable. Turbidity, total suspended solids (TSS), and color are three closely related but distinct measurements that tell operators how well the plant is removing particles and dissolved matter. Regulators set limits on these parameters because high turbidity can shield pathogens from disinfection, making it a direct public-health concern. Understanding how to measure each parameter accurately—and what the numbers mean—is a core operator skill.
Turbidity
Turbidity is a measure of how much light is scattered by particles suspended in water. It is expressed in Nephelometric Turbidity Units (NTU). The measurement is made with a nephelometer (also called a turbidimeter), which shines a beam of light through the sample and detects light scattered at a 90-degree angle to the beam. The more particles present, the more light is scattered, and the higher the NTU reading.
Turbidity is not the same as TSS. A sample can have high turbidity from very fine colloidal particles that pass through a filter, while TSS captures only the particles retained on the filter.
Calibrating and Using a Bench-Top Nephelometer
1. Allow the instrument to warm up per the manufacturer's instructions.
2. Calibrate using traceable standards—either formazin primary standards or commercially prepared StablCal® standards—at the ranges you expect to measure. Always start with the lowest standard and work upward.
3. Wipe the sample cell with a lint-free cloth to remove fingerprints; oils and scratches scatter light and cause false high readings.
4. Fill the cell to the marked line, cap it, and insert it in the same orientation each time (align the index mark).
5. Record the reading in NTU after the display stabilizes.
Acceptable accuracy at the low range (≤ 1 NTU) is critical because drinking-water regulatory limits are set in that range. A reading within ±0.5 NTU of the known standard value is the target for this lab.
Regulatory Limits for Filtered Public Water Supplies
The Safe Drinking Water Act directs the U.S. Environmental Protection Agency (EPA) to set maximum contaminant levels (MCLs) and treatment technique requirements. For filtered surface water systems operating under the Surface Water Treatment Rules, turbidity limits apply at the filter effluent. The specific numeric limits are found in federal and state regulations; your regulatory reference handout contains the current values your state has adopted. As a general principle, turbidity in filtered drinking water must be kept very low—well under 1 NTU in most cases—because:
- Particles can physically shield Giardia cysts, Cryptosporidium oocysts, and bacteria from UV light and chlorine disinfection.
- High turbidity signals that filtration may be failing, prompting immediate investigation.
- Consumers associate turbid water with contamination, creating public-trust issues even when water is technically safe.
Operators must monitor filter effluent turbidity continuously and log exceedances. Surpassing a regulatory trigger value requires specific follow-up actions defined in your state rules.
Total Suspended Solids (TSS) by Gravimetric Method
TSS measures the dry mass of particles retained on a standard glass-fiber filter, reported in milligrams per liter (mg/L). This is a gravimetric method because it depends on weighing—not on optical properties.
Procedure Overview (Standard Methods 2540 D)
1. Prepare the filter. Use a glass-fiber filter with the appropriate pore size (nominally 0.45 µm or as specified). Place it in an aluminum pan and dry it in an oven at 103–105 °C for at least one hour. Cool in a desiccator, then weigh on an analytical balance. Record this as Tare Weight (W1).
2. Filter the sample. Measure a known volume of well-mixed sample (e.g., 100 mL or 1,000 mL depending on expected solids concentration) through the prepared filter under vacuum. Rinse the graduated cylinder with a small amount of distilled water and pass it through the filter.
3. Dry the filter. Transfer the filter and retained solids to the oven at 103–105 °C for at least one hour. Cool in a desiccator.
4. Weigh. Record this as Final Weight (W2).
5. Calculate TSS. See the worked example below.
Avoiding errors: do not let the filter dry out during filtration (causes cracking), do not touch the filter with bare hands (oils add mass), and ensure the oven temperature does not exceed 105 °C (volatile solids may be lost at higher temperatures, biasing the result low).
Color
True color is measured after filtering out turbidity and is caused by dissolved organic matter (humic and fulvic acids from decaying vegetation), iron, or manganese. It is expressed in Platinum-Cobalt Units (PCU) or Color Units (CU). High color can indicate organic precursors that react with chlorine to form disinfection byproducts (DBPs) such as trihalomethanes. Color monitoring guides coagulation chemical doses.
Apparent color includes both dissolved and suspended contributions; true color is measured on filtered samples. This distinction matters when diagnosing treatment problems.
Safety in the Lab
- Safety glasses are required whenever you handle water samples, standards, or chemical reagents.
- Gloves protect against unknown sample constituents; wastewater samples may contain pathogens.
- Formazin stock solution is a suspected carcinogen—work in a ventilated area and dispose of according to your facility's chemical waste procedures.
- The drying oven reaches over 100 °C; use insulated gloves or tongs when removing items. Allow items to cool in the desiccator before handling.
- Wash hands before leaving the lab even if gloves were worn.
- Broken glass-fiber filters or cracked sample cells should be disposed of in a sharps/glass waste container, not the regular trash.
Key terms9
TurbidityThe optical property of water caused by suspended particles scattering light; measured in Nephelometric Turbidity Units (NTU) using a nephelometer.
NTU (Nephelometric Turbidity Unit)The standard unit for reporting turbidity; based on 90-degree light scattering detected by a nephelometer.
Total Suspended Solids (TSS)The mass of particles retained on a glass-fiber filter after drying at 103–105 °C, expressed in milligrams per liter (mg/L); determined by gravimetric analysis.
NephelometerAn instrument that measures turbidity by detecting light scattered at a 90-degree angle from the incident beam; also called a turbidimeter.
Gravimetric MethodAn analytical technique that determines the amount of a substance by precise weighing rather than by optical or chemical reaction.
FormazinA polymer suspension used as a primary turbidity standard because it can be prepared reproducibly; the reference material against which NTU is defined.
True ColorColor measured after filtering out turbidity, caused by dissolved substances such as humic acids, iron, or manganese; reported in Platinum-Cobalt Units (PCU).
Surface Water Treatment Rule (SWTR)A family of EPA regulations that set treatment technique requirements, including turbidity limits at filter effluent, for public water systems using surface water sources.
DesiccatorA sealed container holding a drying agent (desiccant) used to cool filters and other lab items without reabsorbing atmospheric moisture before weighing.
Worked examples4
1. Calculate TSS in mg/L. A 500 mL water sample is filtered through a pre-weighed glass-fiber filter. Tare weight of filter (W1) = 1.5423 g. Weight of filter plus dried solids (W2) = 1.5631 g.
- Step 1 – Find the mass of solids: Mass = W2 − W1 = 1.5631 g − 1.5423 g = 0.0208 g
- Step 2 – Convert grams to milligrams: 0.0208 g × 1,000 mg/g = 20.8 mg
- Step 3 – Convert sample volume to liters: 500 mL ÷ 1,000 mL/L = 0.500 L
- Step 4 – Calculate TSS: TSS = 20.8 mg ÷ 0.500 L = 41.6 mg/L
Answer: TSS = 41.6 mg/L
2. Calculate TSS in mg/L. A 250 mL sample yields a filter tare weight (W1) of 0.8812 g and a final dried weight (W2) of 0.8894 g.
- Step 1 – Mass of solids: 0.8894 g − 0.8812 g = 0.0082 g
- Step 2 – Convert to mg: 0.0082 g × 1,000 = 8.2 mg
- Step 3 – Convert volume to liters: 250 mL ÷ 1,000 = 0.250 L
- Step 4 – TSS = 8.2 mg ÷ 0.250 L = 32.8 mg/L
Answer: TSS = 32.8 mg/L
3. A nephelometer is calibrated with a 10.0 NTU standard. After calibration, the same standard reads 10.3 NTU. A drinking-water filter effluent sample then reads 0.48 NTU. Is the calibration acceptable, and is the sample result within the low-range accuracy target?
- Step 1 – Check calibration error: |10.3 − 10.0| = 0.3 NTU. The lab target for this exercise is ±0.5 NTU at the low range (≤ 1 NTU standards). The 10 NTU check is at a higher range; most instruments specify a percentage error at higher ranges. At 10 NTU, 0.3 NTU represents 3%, which is within a typical ±5% manufacturer specification—so calibration is acceptable.
- Step 2 – Assess the sample reading: 0.48 NTU is below 1 NTU. To confirm low-range accuracy, compare the 1 NTU standard reading to its known value; if that difference is ≤ 0.5 NTU, the result is reliable.
- Step 3 – Interpret the sample: 0.48 NTU is a low turbidity value, consistent with well-treated filtered drinking water meeting regulatory requirements in most states.
Answer: Calibration is acceptable (within manufacturer tolerance); the 0.48 NTU sample result is reliable provided the low-range standard also verified within ±0.5 NTU.
4. A student filters 1,000 mL of a prepared sample with a known TSS of 55.0 mg/L. The student's gravimetric result is 48.5 mg/L. Does this meet the ±10 mg/L accuracy criterion for the lab exercise?
- Step 1 – Calculate the difference: |55.0 − 48.5| = 6.5 mg/L
- Step 2 – Compare to criterion: The acceptable error is ±10 mg/L. 6.5 mg/L < 10 mg/L.
Answer: Yes, 6.5 mg/L difference is within the ±10 mg/L criterion; the result passes.
The math in this lesson1
Learners convert filter mass gain 0.0208 g to mg and 500 mL to L to get TSS 41.6 mg/L.
N-Q.1
In class: Instructor demo: operation and calibration of a bench-top nephelometer using formazin or StablCal standards; students practice calibration and measure at least four samples of different turbidities. · Gravimetric TSS exercise: students filter pre-measured water samples through pre-weighed glass-fiber filters, dry at correct temperature, and calculate TSS in mg/L. · Regulation mapping activity: using a regulatory reference handout, students record the turbidity limits applicable to filtered public water supplies and discuss the health reasons behind the limits in a guided class discussion.
Materials: Benchtop nephelometer, turbidity standards, glass-fiber filters (0.45 µm), filtration apparatus, drying oven, analytical balance, water samples with varying solids, regulatory reference handout, Standard Methods, safety glasses, gloves.
Check: Gravimetric TSS result within ±10 mg/L of known prepared sample; turbidity reading within ±0.5 NTU of standard at low range; student correctly states applicable regulatory limit on an exit slip (80% pass mark across all three checks).