HLT-EKGT89-03 · Lesson 6 · 4 clock hours · theory
Paper Speed, Gain, and Filter Settings
Objective. Students will explain the effect of paper speed, gain, and filter settings on waveform appearance and select appropriate settings for standard adult clinical acquisition.
⚠ Pending professional review
This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.
VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.
Why Machine Settings Matter
A 12-lead EKG is only as useful as the tracing it produces. Before you press Acquire, three machine settings directly control what you see on paper: paper speed, gain (sensitivity), and filters. Choosing the wrong setting can make a normal rhythm look abnormal, hide a true abnormality, or produce a tracing so cluttered with noise that the interpreting clinician cannot read it. Understanding each setting—and knowing the standard values—is a core technician skill and a tested topic on the NHA CET exam.
---
Paper Speed
Paper speed controls how fast the paper moves under the stylus (or how wide each time unit appears on a digital printout). It is measured in millimeters per second (mm/s).
- Standard adult speed: 25 mm/s. At this speed, each small box on the grid equals 0.04 seconds, and each large box (five small boxes) equals 0.20 seconds. This is the universal default for clinical 12-lead EKGs.
- 50 mm/s is sometimes used when the heart rate is very fast (tachycardia) or when the clinician needs to spread waveforms apart to measure intervals more precisely.
Effect on appearance: Increasing speed from 25 to 50 mm/s stretches waveforms horizontally. The same P-QRS-T complex looks wider and easier to measure at 50 mm/s. However, fewer beats fit on a page, and every recording analyzed at a non-standard speed must be clearly labeled, or the reader may misinterpret interval durations. Never submit a 50 mm/s tracing without marking it as such.
---
Gain (Sensitivity/Amplitude)
Gain—also called sensitivity—controls how tall the waveforms appear on paper relative to the actual voltage the heart produces. It is measured in millimeters per millivolt (mm/mV).
- Standard gain: 10 mm/mV. At this setting, 1 millivolt of cardiac electrical signal produces a deflection exactly 10 mm tall. The calibration pulse at the start of each tracing should stand exactly 10 mm high.
- 5 mm/mV (half standard) is used when QRS complexes are so tall that they run into adjacent leads or off the paper—common in athletes and patients with left ventricular hypertrophy.
- 20 mm/mV (double standard) is used when waveforms are very small (low voltage), such as in some patients with pericardial effusion or obesity, to make tiny deflections visible.
Effect on appearance: Doubling the gain doubles every waveform's height. Cutting gain in half shrinks every waveform by half. The rhythm and intervals are unchanged; only the visual size changes. Always note any non-standard gain on the printout and in your documentation so the reader knows the tracing has been scaled.
Safety point: A tracing acquired at double gain can falsely suggest ST elevation or voltage criteria for hypertrophy. A tracing at half gain can make true ST elevation appear normal. Always verify the calibration pulse before handing any tracing to a clinician.
---
Filter Settings
Filters reduce electrical interference that appears as noise on the baseline. Two filters are commonly available:
- Muscle (EMG) filter / high-frequency filter: Reduces high-frequency interference caused by patient movement, shivering, or muscle tremor. Turning this filter on smooths the baseline but may slightly alter the true shape of high-frequency components such as the terminal portion of QRS. Use it when artifact is preventing a readable tracing—but note it on the record.
- Baseline wander (AC / low-frequency) filter: Reduces slow, rolling baseline drift caused by respiration, poor electrode contact, or patient movement. Correcting the root cause (re-preparing the skin, repositioning electrodes, asking the patient to lie still and breathe normally) is always preferred before engaging this filter.
- AC (60 Hz) filter: Removes electrical interference from nearby equipment or power lines.
Standard practice: Acquire the first tracing with the default factory filter settings (usually AC filter on, others off). If artifact persists after correcting technique, engage additional filters and document that you did so. Filters should supplement good technique, not replace it.
---
Selecting Settings for Standard Adult Clinical Acquisition
For a routine adult 12-lead EKG, the correct starting settings are:
| Setting | Standard Value |
|---|---|
| Paper speed | 25 mm/s |
| Gain | 10 mm/mV |
| AC filter | On |
| Other filters | Off unless artifact persists |
Depart from these defaults only when there is a clinical reason. Document every non-standard setting directly on the tracing before it leaves your hands.
---
Scope of Practice and Safety Reminders
- Your role is acquisition, not interpretation. You select settings to produce a technically correct tracing; the clinician interprets what the heart is doing.
- Always check the calibration pulse before and after acquisition. If it does not measure 10 mm at standard gain, the machine needs service.
- If a setting change produces a tracing that looks dramatically different from a previous one for the same patient, flag it for the clinician rather than filing it silently.
- Never alter a printed tracing or its labels after the fact to correct a documentation error. Note the error and re-acquire if clinically indicated.
Key terms8
Paper speedThe rate at which EKG paper moves, measured in mm/s; 25 mm/s is the standard adult setting and determines how wide each time interval appears on the tracing.
Gain (sensitivity)The amplification of the cardiac electrical signal expressed in mm/mV; standard is 10 mm/mV, meaning 1 mV produces a 10 mm deflection on paper.
Calibration pulseA standardized square wave printed at the start of a tracing to confirm that gain is set correctly; at 10 mm/mV, it should measure exactly 10 mm tall.
Baseline wanderA slow, rolling movement of the isoelectric line caused by respiration, poor electrode contact, or patient movement; addressed first by correcting technique, then by filter if needed.
AC (60 Hz) filterA filter that eliminates interference from alternating-current electrical equipment and power lines, reducing rapid regular noise on the tracing.
Muscle artifact filterA high-frequency filter that reduces noise caused by skeletal muscle activity such as shivering or tremor; may slightly alter true high-frequency waveform components.
Half standard (5 mm/mV)A gain setting that reduces waveform height by half, used when large-amplitude QRS complexes extend beyond the printable area of the grid.
Isoelectric lineThe flat baseline on an EKG representing the period when no net electrical activity is detected; wander or noise on this line can distort waveform measurement.
Worked examples4
1. A tracing is acquired at 25 mm/s. The QRS duration measures 5 small boxes wide. How many seconds does that represent, and how wide would the same QRS appear (in small boxes) if the speed were changed to 50 mm/s?
- At 25 mm/s, each small box = 0.04 seconds.
- QRS duration = 5 boxes × 0.04 s/box = 0.20 seconds.
- At 50 mm/s the paper moves twice as fast, so each small box still represents the same voltage but now covers half the time: each small box = 0.02 seconds.
- Same 0.20-second QRS at 50 mm/s: 0.20 s ÷ 0.02 s/box = 10 small boxes wide.
Answer: At 25 mm/s the QRS is 0.20 seconds (5 small boxes). At 50 mm/s the same QRS appears 10 small boxes wide because the paper moves twice as fast, stretching the waveform horizontally.
2. A calibration pulse measures 5 mm tall on a printout labeled with no special notation. What gain was the machine set to, and what is the standard gain the machine should have used?
- Standard gain: 10 mm/mV produces a 10 mm calibration pulse.
- The pulse measured 5 mm, which is exactly half of 10 mm.
- Half the standard deflection corresponds to half the standard gain: 5 mm/mV.
- The machine was set to 5 mm/mV (half standard) instead of 10 mm/mV.
Answer: The machine was set to 5 mm/mV (half standard gain). The correct standard setting is 10 mm/mV, which produces a 10 mm calibration pulse.
3. At standard gain (10 mm/mV), a patient's R wave in lead V5 measures 28 mm tall. If the technician switches to half gain (5 mm/mV) to prevent the complex from running off the grid, how tall will that same R wave appear on the new tracing?
- Original height at 10 mm/mV: 28 mm.
- Halving the gain to 5 mm/mV reduces every waveform's printed height by half.
- New printed height = 28 mm ÷ 2 = 14 mm.
- The actual cardiac voltage is unchanged; only the visual representation is scaled down.
Answer: At 5 mm/mV the R wave will appear 14 mm tall. The clinician must be informed of the non-standard gain so voltage measurements are not underestimated.
4. A technician needs to calculate what voltage a 22 mm deflection represents at standard gain (10 mm/mV). What is the voltage?
- At 10 mm/mV, 10 mm of deflection = 1 mV.
- Set up the proportion: 22 mm ÷ 10 mm/mV = 2.2 mV.
- The deflection represents 2.2 mV of cardiac electrical activity.
Answer: A 22 mm deflection at 10 mm/mV represents 2.2 mV.
In class: Examine printed tracing pairs showing the same rhythm recorded at 25 mm/s vs. 50 mm/s and at 5 mm/mV vs. 10 mm/mV gain, then write a brief explanation of how each change alters waveform appearance. · Complete a worksheet matching five clinical scenarios (large-amplitude QRS, tachycardia, baseline drift) to the recommended setting adjustment and clinical rationale. · Simulate a setting error in class—acquire a tracing at incorrect gain, identify the problem on the printout, correct the setting, and re-acquire, documenting both tracings with a note explaining the change.
Materials: Printed tracing comparison sets, EKG machine, settings worksheet, NHA CET exam content outline
Check: Settings scenario worksheet scored by instructor; students must correctly match settings to scenarios and provide rationale with ≥80% accuracy (4 of 5 scenarios correct).
Practice check8