Digital Imaging Systems, PACS, and Image Post-Processing

Objective. Students explain how computed radiography (CR), digital radiography (DR), and flat-panel detector systems acquire and process images, and perform basic post-processing and PACS workflow tasks on the lab workstation.

⚠ Pending professional review

This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.

VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.

Digital Imaging Systems in Radiologic Technology

Modern radiologic practice relies almost entirely on digital image acquisition. Unlike conventional film-screen radiography, digital systems capture x-ray energy electronically, allowing images to be displayed, adjusted, stored, and transmitted without chemical processing. Two broad categories of digital imaging are used in clinical practice: computed radiography (CR) and digital radiography (DR).

Computed Radiography (CR)

CR uses a flexible photostimulable phosphor (PSP) plate housed in a cassette that resembles a traditional film cassette. When exposed to x-rays, the phosphor crystals store a latent image as trapped electrons in excited energy states. The cassette is then fed into a CR reader, where a laser beam scans the plate and causes the stored electrons to release their energy as visible light — a process called photostimulated luminescence (PSL). A photomultiplier tube detects this light and converts it to a digital signal.

After readout, the plate is exposed to bright white light to erase any residual image so it can be reused. Failure to erase a plate fully can leave a ghost image (erasure artifact) on the next exposure. CR plates must be protected from background radiation and used within a reasonable time after exposure, because the latent image fades.

CR is considered an indirect capture system and requires more steps than DR, but its cassette-based design makes it compatible with existing x-ray rooms and portable equipment.

Digital Radiography (DR)

DR systems capture the image directly in the detector panel, eliminating the cassette-to-reader step. Most modern DR units use a flat-panel detector (FPD) containing a large array of thin-film transistors (TFTs). Two main designs exist:

DR images appear on the monitor within seconds of exposure, improving workflow efficiency. Because there is no cassette exchange, positioning errors are identified faster, reducing patient exposure from repeat examinations.

Exposure Indicators

All digital systems use an exposure indicator (EI) — a numeric value calculated from the detector signal — to report whether the receptor received an appropriate amount of radiation. The International Electrotechnical Commission (IEC) established a standardized EI system with a target EI and a deviation index (DI) to compare the actual exposure to the target. A DI near zero indicates proper exposure. A significantly positive DI suggests overexposure; a significantly negative DI suggests underexposure.

Because the digital system can brighten an underexposed image through post-processing, the technologist must not rely on image appearance alone to judge exposure adequacy. Always check the exposure indicator against the department's reference range. Repeated overexposure increases patient dose unnecessarily — a direct patient safety concern — while underexposure introduces quantum noise and may compromise diagnostic quality.

Image Post-Processing

Once acquired, digital images can be adjusted before or after transmission. Key post-processing tools include:

PACS and RIS Workflow

A Picture Archiving and Communication System (PACS) stores, retrieves, and transmits digital images across a network. It works alongside a Radiology Information System (RIS), which manages patient scheduling, orders, and reports. Together they replace film libraries and paper orders.

A typical technologist workflow includes: (1) verify patient identity using at least two identifiers; (2) confirm the exam order in the RIS; (3) perform the examination; (4) flag the acquired images in PACS for quality review; (5) apply appropriate W/L and annotations; (6) route the completed study to the radiologist's reading queue. Each step must follow department protocol to maintain image integrity and patient privacy under applicable health information regulations.

Digital Image Artifacts

Common digital artifacts include:

Identifying artifacts accurately is essential. A misidentified artifact may lead to a repeated exposure, increasing patient dose, or to a missed pathology if an artifact mimics anatomy.

Safety Considerations

Digital imaging does not eliminate radiation risk. The technologist remains responsible for applying ALARA (As Low As Reasonably Achievable) principles, accurate collimation, appropriate technique selection, and shielding where indicated. Overexposure is easier to commit in digital imaging because a poor technique may still produce a visually acceptable image — this phenomenon is sometimes called exposure creep. Regular audit of exposure indicators and adherence to technique charts are the primary safeguards.

Key terms10

Photostimulable Phosphor (PSP) PlateThe image receptor used in CR; stores x-ray energy as trapped electrons that are released as light when scanned by a laser during readout.
Flat-Panel Detector (FPD)The solid-state image receptor used in DR systems; contains a TFT array that converts x-ray energy to a digital signal, either directly or via a scintillator layer.
Exposure Indicator (EI)A numeric value generated by a digital imaging system that reflects the amount of radiation received by the detector, used to assess whether technique was appropriate.
Deviation Index (DI)A standardized IEC metric comparing the actual exposure indicator to the target EI; values near zero indicate proper exposure, positive values indicate overexposure, negative values indicate underexposure.
Windowing and Leveling (W/L)Post-processing adjustment of image contrast (window width) and brightness (window level/center) to optimize visualization of specific tissue types without repeating the exposure.
PACS (Picture Archiving and Communication System)A networked system that stores, retrieves, and transmits digital radiographic images throughout a healthcare facility and to remote locations.
Quantum Noise (Mottle)A grainy image appearance caused by an insufficient number of x-ray photons reaching the detector; associated with underexposure and a low exposure indicator.
Exposure CreepThe gradual tendency to increase patient exposure over time in digital imaging because the system compensates for poor technique and still produces a visually acceptable image.
Photostimulated Luminescence (PSL)The process by which a CR plate releases stored x-ray energy as visible light when stimulated by a laser beam during readout in the CR reader.
RIS (Radiology Information System)A database system used in radiology departments to manage patient orders, scheduling, demographic data, and report generation; works in concert with PACS.
In class: Complete a structured comparison exercise: for CR and DR systems, identify the detector type, image receptor, readout process, exposure indicator system, and primary advantage or limitation, presented as a completed two-column table submitted for grading. · At the PACS workstation, practice the full workflow for five simulated cases — patient registration, exam order entry, image acquisition flag, windowing and leveling optimization, annotation, and electronic send to radiologist queue — following department protocol. · Analyze a set of 10 digital images for artifacts (grid lines, erasure artifacts, quantum noise, detector damage) and for over- or under-exposure as evidenced by exposure indicator values outside the reference range, documenting findings on a structured artifact log.
Materials: CR reader and phosphor plates, DR flat-panel unit, PACS/RIS simulation software or live department workstation, exposure indicator reference chart, artifact log worksheet, Bushong's Radiologic Science for Technologists (current edition)
Check: Comparison table graded for accuracy (80% correct); PACS workflow simulation must be completed error-free for at least 4 of 5 cases; artifact log must correctly identify artifact type and probable cause for 8 of 10 images — pass 80%.

Practice check8

Answer each question, then check it. Nothing is recorded; this is your own practice.

Image Production
1. In computed radiography, what process releases the stored latent image from the photostimulable phosphor plate during readout?
Image Production
2. A technologist notices a faint outline of a previous patient's chest on a new abdominal image acquired with a CR system. What is the most likely cause?
Image Production
3. Which flat-panel detector design converts x-ray energy directly to an electrical charge without an intermediate light step?
Radiation Protection
4. A deviation index (DI) of +3 on a digital image most likely indicates which of the following?
Image Production
5. A radiograph appears grainy and the exposure indicator is well below the department's reference range. What is the most appropriate initial action?
Patient Care and Education
6. Which PACS workflow step is performed first to ensure images are linked to the correct patient record?
Image Production
7. Adjusting the window width on a digital chest radiograph will primarily change which image characteristic?
Radiation Protection
8. A technologist consistently uses higher-than-recommended mAs values because digital images always look acceptable on the monitor. This practice is best described as which patient safety concern?