Objective. Students produce diagnostically acceptable periapical radiographs of all regions of the mouth using the paralleling technique with film-holding instruments on a typodont.
This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.
VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.
The paralleling technique is the preferred method for taking periapical radiographs because it produces images with the least geometric distortion. The core principle is simple: the image receptor (film, PSP plate, or digital sensor) is placed parallel to the long axis of the tooth, and the central ray of the x-ray beam is directed perpendicular (at a right angle) to both the receptor and the tooth. This geometry minimizes elongation and foreshortening, giving the dentist an accurate picture of tooth length, root morphology, and surrounding bone.
Compare this to the older bisecting-angle technique, which angles the beam toward an imaginary line between the tooth and the receptor. The bisecting technique is more prone to distortion and is used only when the paralleling technique is not possible.
Because the receptor must be held away from the tooth (often near the center of the mouth), patients cannot simply bite on it. Standardized film holders — commonly the Rinn XCP (Extension Cone Paralleling) system or equivalent — hold the receptor in the correct position and attach to an aiming ring and indicator rod that guide the tube head to the right angulation. Each component has a role:
XCP kits include color-coded components for anterior (blue) and posterior (yellow) regions. Some manufacturers use a third set (red) for bitewing radiographs. Learn your clinic's specific system.
A standard full-mouth series (FMS) consists of 18 periapical images covering every region of the dentition. Placement guidelines:
Always place the receptor so the active surface faces the tube head. With PSP plates, the white side faces the beam; with sensors, follow the manufacturer's orientation mark. Reversed placement causes a severe underexposure with a herringbone artifact (for film/PSP) or a blank/reversed image (for sensors).
Once the holder is in place, slide the PID until it rests against the aiming ring. The ring ensures:
1. Correct horizontal angulation — the beam passes through the contact areas of the teeth, avoiding overlapping
2. Correct vertical angulation — the beam is perpendicular to both the receptor and the tooth, preventing elongation or foreshortening
3. Correct centering — the beam covers the receptor fully, preventing cone cuts
Never guess the angulation freehand when using an XCP system. Trust the ring.
| Error | Cause | Correction |
|---|---|---|
| Elongation | Vertical angulation too flat (not enough) | Increase vertical angulation |
| Foreshortening | Vertical angulation too steep (too much) | Decrease vertical angulation |
| Cone cut (clear unexposed area) | PID not centered on aiming ring | Re-align PID flush with ring |
| Overlapping contacts | Incorrect horizontal angulation | Adjust PID horizontally through contacts |
| Blurry image | Patient or receptor movement | Instruct patient to hold still; secure holder |
| Light image (underexposed) | Low kVp/mA, short time, or reversed receptor | Check unit settings; correct receptor orientation |
| Dark image (overexposed) | High exposure settings | Reduce exposure factors |
Every exposure, even on a phantom, requires the same safety habits you will use in clinic.
The DANB Radiation Health and Safety (RHS) component exam — one of the exams that make up the CDA credential — tests your knowledge of paralleling technique, receptor placement, error identification, and radiation protection. Your state dental radiology license exam tests similar content with a legal emphasis: operating an x-ray unit without a license is a violation of state law. Mastery of paralleling technique is therefore both a clinical skill and an exam objective.
Answer each question, then check it. Nothing is recorded; this is your own practice.
A free tech program — complimentary, as our commitment to get America working. Help us build VoTechWorks. Tell us what works, what doesn’t, and what you want to learn.
Does the school offer financial aid?
No. We don’t offer financial aid because we’re free. No, really: completely free.
Do you offer industry certification?
No. We are not certified yet, so we can’t offer that to students. What we do offer is knowledge and understanding of a trade.
Enroll first in the trade you’re interested in. You get your key, your program opens, and you begin studying.
Enroll free →