Objective. Students will select the appropriate packaging material for each instrument set, wrap trays using sequential (double) wrap technique, seal pouches correctly, and apply all required labeling elements.
This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.
VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.
Sterilization is only as good as the package that maintains sterility until the moment of use. A package protects the sterilized item from microbial contamination, moisture, and physical damage during storage and transport. If the package is compromised—by a pinhole, an incomplete seal, or improper handling—the contents must be considered contaminated and cannot be used on a patient. Selecting the right packaging material and applying correct technique are therefore direct patient safety responsibilities of the surgical technologist.
SMS Wrap (Spunbond-Meltblown-Spunbond)
SMS wrap is a nonwoven fabric made of polypropylene in three bonded layers. It is the most common reusable-cycle wrap for wrapping instrument trays. SMS wrap is a sterilant-penetrable barrier that also repels liquids after sterilization. It is available in different weights (commonly expressed in grams per square meter). Single-use SMS wrap is the current standard; reusable woven textile wraps are rarely seen in modern facilities. Always check the manufacturer's instructions for the correct weight and size for your tray.
Peel Pouches
Peel pouches are used for individual instruments or small items. Two common types are:
Pouches must be sized so that the item fits without stretching or stressing the seams. Leave approximately one inch of space beyond the item toward the seal end. Double-pouching (placing one sealed pouch inside a second pouch, paper to plastic) is acceptable when required by the procedure but must follow facility policy.
Rigid Sterilization Containers
Rigid containers are reusable aluminum or stainless-steel systems with filtered or valve-controlled lids. Filters (disposable) or valves allow sterilant to penetrate during the cycle and then act as a barrier after sterilization. Rigid containers are validated by the manufacturer for specific sterilizers and cycle parameters. After every use, inspect gaskets, locks, and filters. Replace filters and any damaged components before reprocessing.
Sequential wrapping places two individual sheets of SMS wrap around the tray one after the other, each folded completely before the next is applied. This provides a two-layer barrier and allows aseptic presentation (opening) at the sterile field.
Envelope Fold (Diagonal)
1. Place the tray in the center of the wrap sheet, rotated 45° so one corner points toward you.
2. Fold the near corner up and over the tray, tucking it back slightly.
3. Fold the right corner in, then the left corner in, overlapping.
4. Fold the remaining (far) corner up and over; the tip folds back to form a cuff that can be grasped during opening without breaking sterility.
5. Secure with sterilization indicator tape.
6. Repeat with a second sheet in the same manner.
Square (Parallel) Fold
1. Place the tray in the center with sides parallel to the wrap edges.
2. Fold the near edge up and over, creasing tightly.
3. Fold the far edge over.
4. Fold the right side in, then the left side over it.
5. Secure with sterilization indicator tape.
6. Repeat with a second sheet.
Regardless of fold style, folds must be tight with no gaps or loose flaps. Gaps allow microorganisms to enter.
Heat-seal pouches using a validated heat sealer. The seal must be complete across the full width of the pouch with no channels or wrinkles. After sealing, inspect the seal by gently pressing edges—a proper seal shows no separation. Self-sealing (peel-and-stick) pouches are folded and pressed firmly; check the manufacturer's instructions for required pressure and technique. Never use staples, pins, or tape on peel pouches, as these breach the barrier.
Every sterile package must be labeled before sterilization or immediately after, according to facility policy. Required elements (per AAMI ST79 guidance and standard accreditation requirements) include:
Write or apply labels on the indicator tape or a dedicated label area—never on the paper side of a peel pouch in a way that damages the barrier.
The current standard supported by AAMI ST79 recognizes that sterility is maintained until an event compromises the package (a tear, puncture, wetting, or opened seal), not simply because a calendar date has passed. This is called event-related sterility. Facilities may still assign maximum expiration dates based on storage conditions and package type as part of their policy. Heavy-traffic storage, open shelving, and handling all increase the risk of contamination. Rigid containers in closed, limited-access storage generally maintain sterility longer than wrapped trays on open shelves.
Always inspect every package before opening: look for seal integrity, moisture, tears, and a changed (failed) chemical indicator. If any doubt exists, do not use the package.