Objective. Students will perform point-of-use treatment, manual scrubbing with enzymatic detergent, and ultrasonic cleaning of soiled surgical instruments following AAMI ST79 guidelines.
This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.
VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.
Every instrument that touches a patient carries blood, tissue, and body fluids. Before sterilization or high-level disinfection can work, all organic material and debris must be physically removed. Sterilants and disinfectants cannot penetrate a biofilm or soil layer, so an instrument that looks clean but still carries residue will not be reliably sterile after processing. Decontamination — the combination of cleaning and disinfection steps that reduce the microbial load to a safe level — is therefore the most critical phase of instrument reprocessing. AAMI ST79 (the Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities) is the primary US guidance document that describes best practice for every step covered in this lesson.
Point-of-use treatment begins at the sterile field, the moment an instrument is no longer needed. The scrub surgical technologist removes gross soil — visible blood, tissue, and saline — by wiping instruments with a sterile, moist sponge and rinsing them in sterile water when available. The goal is to prevent soil from drying and hardening, which makes later cleaning far more difficult. Instruments must never be allowed to dry before reprocessing.
After the procedure, used instruments are placed in a covered, puncture-resistant container (often a closed rigid tray or a covered basin) before transport to the decontamination area. This protects both the instruments and the personnel who carry them. Sharps — including needle holders with suture needles — must be accounted for and contained safely before the set leaves the room.
The decontamination area is a restricted zone with a defined PPE requirement. Before touching any soiled instrument, personnel must don:
Removing PPE in the wrong order or leaving the decontamination area without removing soiled PPE risks spreading contamination. This is a patient-safety and worker-safety issue.
Manual cleaning follows these key principles.
Water temperature. Use water at the temperature specified in the detergent's instructions for use (IFU). Most enzymatic detergents are formulated for lukewarm water — typically around 100–110 °F (roughly 38–43 °C). Water that is too hot denatures the enzymes and destroys cleaning effectiveness; water that is too cold slows enzymatic activity.
Dilution. Mix detergent to the exact ratio stated in the IFU. More is not better: excessive concentration can leave residue that interferes with sterilization.
Brush technique. Use an appropriately sized soft-bristle brush. Scrub all surfaces under the waterline to prevent aerosol generation. Open box-lock joints fully. Scrub inside serrations, lumens, and crevices. Use pipe cleaners or lumen brushes for tubular instruments.
Rinsing. Rinse instruments thoroughly with water — the quality of rinse water matters. Hard water can leave mineral deposits. A final rinse with treated (softened or deionized) water is recommended when available.
Inspection. After rinsing, visually and functionally inspect each instrument: check that box locks move freely, scissors cut cleanly, and ratchets engage. Damaged or permanently soiled instruments are removed from service.
Document the process on a decontamination log that records the date, instrument set, operator, water temperature, detergent lot, and any instruments removed.
Ultrasonic cleaners use a process called cavitation — the rapid formation and implosion of microscopic bubbles created by high-frequency sound waves — to dislodge soil from instrument surfaces, including areas a brush cannot reach.
Loading the ultrasonic unit. Instruments must be:
Do not mix metals (e.g., stainless steel with chrome-plated instruments) in the same load; dissimilar metals can cause electrolytic damage.
Degassing cycle. A new batch of solution contains dissolved air that inhibits cavitation. Run a degassing cycle — typically a short timed run with the lid open — before the first cleaning cycle of the day or after each solution change.
Cleaning cycle. Run the cycle for the time specified in the manufacturer's IFU. Extend the cycle only if the IFU allows; over-exposure can pit instrument surfaces.
Cavitation test (aluminum foil test). Place a small piece of aluminum foil in the energized unit for approximately 30–60 seconds per the manufacturer's IFU. Pitting and perforation of the foil indicate that cavitation is occurring. Smooth or undamaged foil signals inadequate cavitation — stop and troubleshoot before using the unit.
After ultrasonic cleaning. Remove instruments, rinse thoroughly, inspect, and lubricate hinged instruments with an instrument milk (water-soluble lubricant compatible with sterilization) before assembly and packaging.