HLT-STER66-02 · Lesson 4 · 18 clock hours · lab

Manual Cleaning: Techniques, Chemistry, and Instrument Disassembly

Objective. Students will manually clean a set of at least eight surgical instruments—including box-lock, lumen, and multi-part items—following the IFU for each instrument and the selected cleaning chemistry, achieving visually soil-free results on post-cleaning inspection.

⚠ Pending professional review

This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.

VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.

Why Manual Cleaning Is the Foundation of Sterile Processing

No sterilization method—steam, ethylene oxide, or low-temperature plasma—can penetrate and destroy pathogens that are shielded inside dried blood, tissue, or biofilm. Manual cleaning removes that organic burden before any disinfection or sterilization step begins. If cleaning fails, sterilization fails, and patients are at risk. For this reason, manual cleaning is considered the most critical step in the entire sterile processing cycle.

Reading the Instrument Manufacturer's IFU First

Before you touch an instrument with water or detergent, you must locate and read the manufacturer's Instructions for Use (IFU). The IFU is a legally required document that specifies exactly how that instrument must be cleaned, disassembled, brushed, rinsed, and dried. Key details to record from every IFU include:

IFU binders and laminated IFU cards should be kept at the decontamination sink and consulted every shift, not memorized and assumed.

Personal Protective Equipment in the Decontamination Area

All work in decontamination is performed in a full PPE ensemble: fluid-resistant gown, surgical mask or face shield, heavy-duty puncture-resistant gloves (not exam gloves), and shoe covers. Eye protection is mandatory—aerosolized contaminated water is an exposure risk. PPE is donned before entering the decontamination area and removed before leaving. Never carry PPE into clean areas.

Disassembly: Getting to Every Surface

Instruments that are not fully disassembled before cleaning cannot be cleaned effectively. Rules for disassembly:

Place all components together as a set so nothing is lost during cleaning.

Cleaning Chemistry: Enzymatic vs. Neutral pH Detergent

Enzymatic detergents contain enzymes (typically protease, lipase, and amylase) that actively break down proteins, fats, and carbohydrates in organic soil. They work best at the temperature specified by their IFU and lose activity if the solution is too hot, too dilute, or too concentrated. Enzymatic solutions are single-use: discard after each sink load or per manufacturer guidance.

Neutral pH detergents (pH approximately 6–8) are general-purpose cleaners that do not contain active enzymes. They are appropriate for routine cleaning of instruments that do not carry heavy organic soil, and for facilities where the IFU permits their use.

Neither type should be mixed together unless the product manufacturer specifically permits it. Always dilute to the manufacturer-specified concentration—more is not better. Over-concentrated detergent leaves residue that can interfere with sterilization.

The Wash-Rinse-Rinse Sequence

Manual cleaning follows a structured sequence:

1. Pre-rinse at point of use (done in the OR or procedure area before transport): removes gross soil and prevents drying.

2. Wash step: submerge instruments in detergent solution, brush all surfaces systematically—serrations, box-locks, hinges, lumens, and exterior surfaces. Keep instruments below the waterline while brushing to prevent aerosolization of contaminated droplets.

3. First rinse: remove all detergent and loosened soil with tap water.

4. Final rinse: use purified water (deionized or distilled) if the IFU requires it to prevent mineral spotting.

Maintain water levels to the fill lines marked on the sink. Overfilling increases splash risk; underfilling reduces the effectiveness of the soak.

Brushing Technique

Select the brush size that fits the lumen or feature being cleaned—it should contact the inner wall with slight resistance. Use a back-and-forth motion through lumens so that both ends of the brush exit the channel, ensuring the entire length is scrubbed. For serrations and jaw surfaces, angle the brush into the teeth. For box-locks, work the brush into the joint with the instrument held open. Inspect brush bristles before use; a bent or worn brush does not clean effectively and must be discarded.

Post-Cleaning Inspection

After the final rinse, lay instruments on a clean towel. Under a magnifying glass and inspection light, check:

If any soil is visible, the instrument must be recleaned. Document each cleaning attempt in the lab log, including detergent lot number, dilution used, water temperature, and inspector initials. This documentation habit mirrors the quality records required in every accredited sterile processing department.

Scope of Practice and Safety Reminders

Key terms10

IFU (Instructions for Use)Manufacturer-written document specifying the exact steps required to safely clean, disassemble, disinfect, and sterilize a device; legally required to be followed and supersedes general department protocols.
Enzymatic detergentA cleaning solution containing active enzymes (such as protease, lipase, and amylase) that break down organic soil including proteins, fats, and carbohydrates; single-use and temperature-sensitive.
Neutral pH detergentA general-purpose instrument cleaning solution with a pH of approximately 6–8 that does not contain active enzymes; used when IFU permits and heavy organic soil is not present.
Box-lockThe pivot joint of hinged surgical instruments such as hemostats and needle holders; must be fully opened during cleaning to allow brush and detergent access to the joint area.
BiofilmA structured community of microorganisms encased in a self-produced matrix that adheres to instrument surfaces; highly resistant to disinfection and sterilization and prevented by timely, thorough cleaning.
LumenThe hollow interior channel of tubular instruments such as trocars, cannulas, and suction tips; requires a correctly sized brush to clean the entire inner surface.
Dilution ratioThe manufacturer-specified proportion of concentrated detergent to water needed to produce a working-strength cleaning solution; must be measured accurately to ensure efficacy and prevent residue.
AerosolizationThe generation of fine airborne droplets containing contaminated liquid, which can occur when instruments are brushed above the waterline; a key reason all brushing must be done submerged.
DecontaminationThe combination of physical cleaning and, when applicable, disinfection steps used to render a contaminated instrument safe to handle and ready for inspection, assembly, and sterilization.
Dilution logA quality-record document on which a sterile processing technician records the detergent lot number, concentration, water temperature, and date/time for each cleaning solution prepared; supports traceability and compliance.

Worked examples4

1. An enzymatic detergent concentrate requires a dilution of 1 oz of concentrate per 1 gallon of water. How many ounces of concentrate are needed to fill a 3-gallon sink?
  1. Identify the dilution ratio from the IFU: 1 oz concentrate per 1 gallon of water.
  2. Identify the target volume: 3 gallons.
  3. Multiply: 1 oz × 3 gallons = 3 oz of concentrate.
  4. Add 3 oz of concentrate to the sink, then fill to the 3-gallon mark with water at the manufacturer-specified temperature.
Answer: 3 oz of enzymatic concentrate are needed for a 3-gallon sink.
2. A detergent label states: dilute at a ratio of 1:128 (concentrate to water). How many fluid ounces of concentrate are needed to prepare 1 gallon of working solution? (1 gallon = 128 fl oz)
  1. Understand the ratio: 1 part concentrate to 128 parts total solution means 1 fl oz of concentrate per 128 fl oz of finished solution.
  2. Total working solution needed: 1 gallon = 128 fl oz.
  3. Calculate concentrate volume: 128 fl oz ÷ 128 = 1 fl oz of concentrate.
  4. Add 1 fl oz of concentrate to the container, then add water to reach the 128 fl oz (1-gallon) mark.
Answer: 1 fl oz of concentrate is needed to make 1 gallon of working solution at a 1:128 dilution.
3. A decontamination sink holds 5 gallons. The enzymatic detergent IFU specifies a dilution of 1 oz concentrate per 1 gallon of water. After filling, a test strip shows the solution is out of the acceptable concentration range. The technician suspects over-dilution because only 3 oz of concentrate were added. How many additional ounces of concentrate are needed to correct the solution—or should the sink be discarded and remixed?
  1. Determine the correct amount of concentrate for 5 gallons: 1 oz × 5 gallons = 5 oz required.
  2. Determine how much concentrate was actually added: 3 oz.
  3. Calculate the shortage: 5 oz − 3 oz = 2 oz short.
  4. Consider whether adding 2 oz to an already-diluted solution will reach the correct concentration: because enzymatic solutions are sensitive to exact dilution and the solution has already been used, best practice per most manufacturer guidance is to discard the solution and prepare a fresh batch at the correct dilution (5 oz concentrate in 5 gallons of fresh water at the correct temperature).
  5. Document the discard and the new mixture on the dilution log.
Answer: Discard the under-concentrated solution and prepare a fresh 5-gallon batch using 5 oz of concentrate. Document the corrective action on the dilution log.
4. A technician needs to verify that the water temperature in the wash sink is within the range specified by the enzymatic detergent IFU (stated as 'between 100 °F and 115 °F'). The thermometer reads 42 °C. Is this temperature within range? (Use the formula °F = (°C × 9/5) + 32)
  1. Apply the conversion formula: °F = (42 × 9/5) + 32.
  2. Multiply: 42 × 9 = 378; divide: 378 ÷ 5 = 75.6.
  3. Add 32: 75.6 + 32 = 107.6 °F.
  4. Compare to the IFU range of 100 °F–115 °F: 107.6 °F falls within that range.
Answer: 42 °C equals 107.6 °F, which is within the acceptable range of 100 °F–115 °F. The water temperature is acceptable.

The math in this lesson1

Learners scale a 1 oz per gallon detergent dilution to find 3 oz for a 3-gallon sink.
6.RP.3
In class: IFU scavenger hunt: students are assigned four different instruments (e.g., Kelly clamp, laparoscopic trocar, flexible endoscope channel brush, bone rongeur) and must locate and summarize the manufacturer's IFU cleaning steps for each, noting water temperature limits, brush size, detergent type, and disassembly requirements before any water contact. · Hands-on manual cleaning station: working at a sink with fill lines for wash and rinse, students disassemble instruments per IFU, use brushes (correct size for each lumen or serration), enzymatic or neutral pH detergent at manufacturer-specified dilution, and follow a wash-rinse-rinse sequence, then lay cleaned instruments on a towel for inspection by a lab partner using a checklist. · Chemistry dilution math exercise: students calculate the correct volume of concentrated enzymatic cleaner needed to make a working-strength solution in 1-gallon, 3-gallon, and a sink of a specified volume, record on a dilution log, and mix the solution, verifying concentration with a test strip where applicable.
Materials: Decontamination sink with fill lines and splash guard, assorted training or decommissioned instruments (box-lock, lumen types, multi-part), instrument cleaning brushes (multiple sizes), enzymatic detergent concentrate and neutral pH detergent, measuring cups, pH/concentration test strips, IFU binders or laminated IFU cards, magnifying glass, inspection light, PPE ensemble, dilution log forms
Check: Each student's cleaned instrument set is inspected by the instructor under magnification and direct light; pass standard is zero visible soil on all eight instruments and correct documentation of detergent lot, dilution, and water temperature on the dilution log. Any instrument failing inspection must be recleaned and re-inspected before the student may advance (lab log entry required for each attempt).

Practice check8

Answer each question, then check it. Nothing is recorded; this is your own practice.

Decontamination
1. Before beginning manual cleaning of a multi-part laparoscopic instrument, the FIRST action the technician must take is:
Decontamination
2. Why must instruments be brushed while submerged below the waterline during manual cleaning?
Decontamination
3. A technician is cleaning a Kelly clamp and notices the box-lock area still appears brown after the wash step. What is the most likely cause?
Decontamination
4. Which characteristic best describes a properly selected lumen brush for a trocar cannula?
Decontamination
5. After preparing an enzymatic cleaning solution, a technician uses the remaining solution at the end of the shift for a second instrument load. This practice is:
Decontamination
6. A technician prepares a 2-gallon wash solution using a detergent with a 1 oz per gallon dilution ratio but accidentally adds 6 oz of concentrate. What is the likely consequence of using this over-concentrated solution?
Decontamination
7. During post-cleaning inspection, a technician holds a suction tip up to an inspection light and cannot see through the lumen. The correct action is:
Documentation and Record-Keeping
8. Which information is REQUIRED to be documented on a decontamination dilution log?