HLT-RESP80-06 · Lesson 4 · 36 clock hours · practice

Oxygen Delivery Systems: Clinical Selection and Troubleshooting

Objective. Students will select the clinically appropriate oxygen delivery device for a given patient scenario, correctly set up and verify the device, and troubleshoot a minimum of four simulated equipment malfunctions.

⚠ Pending professional review

This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.

VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.

Oxygen Delivery Systems: Clinical Selection and Troubleshooting

Selecting the right oxygen delivery device is one of the most fundamental clinical decisions a respiratory therapist makes. The wrong choice can leave a patient hypoxic, cause carbon dioxide retention in a patient at risk, or deliver far more oxygen than is needed. This lesson covers the six core devices you will use in clinical practice, how to set them up safely, and how to recognize and correct common equipment failures.

The Six Core Devices

Nasal cannula (NC) delivers low-flow oxygen through two soft prongs seated just inside the nostrils. Flow rates typically range from 1–6 L/min, producing an estimated FiO₂ of roughly 0.24–0.44, though the actual FiO₂ varies with the patient's breathing pattern. It is comfortable, allows eating and talking, and is appropriate for patients who need mild supplemental oxygen and can breathe through their nose. Humidification is generally recommended when flow exceeds 4 L/min to prevent mucosal dryness.

Simple mask covers the nose and mouth and contains small side exhalation ports. Flow must be maintained at a minimum of 5 L/min to flush exhaled CO₂ from the mask; flows of 5–10 L/min deliver an estimated FiO₂ of roughly 0.35–0.55. Never run a simple mask below 5 L/min — CO₂ can accumulate inside the mask.

Partial rebreather mask adds a reservoir bag to the simple mask. A portion of the patient's exhaled air — primarily from early exhalation (dead-space gas, which is rich in oxygen and low in CO₂) — refills the bag. Flow must be set high enough that the reservoir bag does not fully collapse on inspiration; typical flows are 6–10 L/min, yielding an estimated FiO₂ up to approximately 0.60.

Non-rebreather mask (NRB) also has a reservoir bag but adds one-way valves: a valve between the bag and mask prevents exhaled gas from entering the bag, and valves over the exhalation ports prevent room air from being inhaled. This produces the highest FiO₂ available from a low-flow system — up to approximately 0.60–0.80 at 10–15 L/min. The bag must remain at least one-third to one-half inflated at all times. If the bag collapses completely, the patient cannot inhale adequate gas — this is a critical malfunction requiring immediate correction.

Venturi mask (air-entrainment mask) is the device of choice when a precise, consistent FiO₂ is required — most importantly in patients at risk for hypercapnic respiratory failure (for example, certain patients with COPD). Color-coded adapters entrain a fixed ratio of room air, producing a specific FiO₂ regardless of the patient's breathing pattern. Common settings range from 0.24 to 0.50. Always confirm the correct color-coded adapter is in place and set the flow rate printed on that adapter.

High-flow nasal cannula (HFNC) delivers heated, humidified oxygen at flows up to 60 L/min, meeting or exceeding the patient's peak inspiratory demand. This allows a more reliable and adjustable FiO₂ (0.21–1.0) and provides a small amount of positive airway pressure. HFNC requires a dedicated blender and heated humidifier system. It is used for moderate-to-severe hypoxemia and patients who need high FiO₂ but are not intubated.

Selecting the Right Device

Ask three questions for every patient:

1. What FiO₂ is needed? Target SpO₂ guides this decision, along with the physician's order and clinical context.

2. Does the patient have hypercapnic risk? If yes, use a Venturi mask to prevent over-oxygenation and unintended suppression of hypoxic drive in susceptible patients.

3. Can the patient cooperate and tolerate the device? An agitated or confused patient may not tolerate a tight-fitting mask; a cannula may be safer even if flow requirements are higher.

Setup and Verification

For every device: (1) confirm the order, (2) assemble all components, (3) connect to a calibrated flow meter, (4) set the prescribed flow rate, (5) check for leaks and proper connections, (6) verify humidification is functioning when required, and (7) apply the device to the patient and reassess SpO₂ within a clinically appropriate interval.

Troubleshooting

| Malfunction | Signs | Corrective Action |

|---|---|---|

| Disconnected tubing | No flow; patient not improving; audible hiss at connection point | Reconnect securely; verify flow resumes |

| Incorrect flow rate | SpO₂ not at target; flow meter ball not at prescribed mark | Adjust flow meter to ordered rate |

| Depleted humidifier water | Gurgling stops; patient reports dryness; water chamber empty | Refill or replace bubble humidifier per facility policy |

| NRB bag not inflating | Bag remains flat or collapses on inhalation | Check that flow is adequate; verify one-way valve is seated correctly; check for leaks at mask seal |

Safety Points

Key terms8

FiO₂ (Fraction of Inspired Oxygen)The proportion of oxygen in the gas a patient inhales, expressed as a decimal from 0.21 (room air) to 1.0 (pure oxygen). Different devices deliver different FiO₂ ranges.
Low-flow oxygen systemA delivery system that supplies oxygen at a flow rate lower than the patient's peak inspiratory demand, so room air is entrained to make up the remainder; actual FiO₂ varies with breathing pattern.
High-flow oxygen systemA delivery system that meets or exceeds the patient's peak inspiratory demand, providing a more consistent and controllable FiO₂; the Venturi mask and HFNC are examples.
Reservoir bagA collapsible bag attached to partial rebreather and non-rebreather masks that stores oxygen between breaths, allowing higher FiO₂ than a simple mask at the same flow.
Venturi (air-entrainment) maskA mask that uses a jet of oxygen through a color-coded adapter to entrain a precise ratio of room air, delivering a specific, reliable FiO₂ independent of the patient's breathing pattern.
Hypercapnic riskThe clinical concern that some patients (for example, those with chronic CO₂ retention) may reduce their respiratory drive if given excess oxygen, causing CO₂ to rise; precise FiO₂ control is essential in these patients.
Bubble humidifierA small water-filled chamber connected between the flow meter and delivery device that adds moisture to oxygen flow, reducing mucosal dryness, especially at higher flow rates.
Non-rebreather mask (NRB)A mask with a reservoir bag and one-way valves that prevents exhaled gas from re-entering the bag and prevents room air entrainment, delivering the highest FiO₂ of any low-flow device.

Worked examples4

1. A patient is on a nasal cannula at 2 L/min. Using the standard estimate (each liter per minute adds approximately 0.04 to a baseline of 0.20), what is the estimated FiO₂?
  1. Start with room air FiO₂: 0.20
  2. Each L/min of flow adds approximately 0.04 to FiO₂
  3. Multiply: 2 L/min × 0.04 = 0.08
  4. Add to baseline: 0.20 + 0.08 = 0.28
Answer: Estimated FiO₂ ≈ 0.28 at 2 L/min nasal cannula. Remember this is an approximation; actual FiO₂ varies with the patient's tidal volume and respiratory rate.
2. A patient is on a nasal cannula at 6 L/min. Using the same estimation formula, what is the estimated FiO₂, and what additional care step is indicated at this flow?
  1. Start with baseline: 0.20
  2. Multiply: 6 L/min × 0.04 = 0.24
  3. Add: 0.20 + 0.24 = 0.44
  4. Note that flow exceeds 4 L/min, so humidification with a bubble humidifier is indicated to prevent nasal mucosal dryness
Answer: Estimated FiO₂ ≈ 0.44; a bubble humidifier should be added because the flow rate exceeds 4 L/min.
3. The ordered FiO₂ via nasal cannula is estimated to be 0.36. At what flow rate should the device be set?
  1. Target FiO₂ = 0.36; baseline = 0.20
  2. Difference: 0.36 − 0.20 = 0.16
  3. Divide by 0.04 per L/min: 0.16 ÷ 0.04 = 4 L/min
Answer: Set the nasal cannula to 4 L/min to achieve an estimated FiO₂ of 0.36.
4. A non-rebreather mask is ordered. Before placing it on the patient, how do you verify the reservoir bag is adequately inflated, and what flow range is appropriate?
  1. Occlude the mask-to-bag valve port with your thumb while gas is flowing to allow the bag to fill completely before application
  2. Observe that the bag inflates fully — it should remain at least one-third to one-half full during the patient's inspiration once applied
  3. Set flow in the range of 10–15 L/min (per facility protocol and physician order) to maintain bag inflation
  4. Confirm no air leaks at the bag seam or mask connections after placing on the patient
Answer: Pre-inflate the bag by briefly occluding the valve before application; set flow at 10–15 L/min and verify the bag does not completely collapse during inhalation.

The math in this lesson1

Learners estimate nasal cannula FiO₂ by adding 0.04 per L/min to the room-air baseline.
F-LE.1
Brush up in the math halls at newSchoolWorks.com:
In class: Given eight patient scenario cards (varying FiO2 need, diagnosis, ability to cooperate, and presence or absence of hypercapnic risk), students select the correct device from a displayed array—nasal cannula, simple mask, partial rebreather, non-rebreather, Venturi mask, high-flow nasal cannula—and justify the choice in writing. · Hands-on setup rotation: each student individually assembles, connects to a flow meter, and verifies flow for five different devices under timed conditions, with instructor observing for correct liter-flow setting, proper humidification when indicated, and correct fit on a manikin head. · Troubleshooting exercise: instructor pre-sets four malfunctions (disconnected tubing, incorrect flow rate, depleted humidifier water, non-rebreather bag not inflating) hidden within assembled devices; students diagnose and correct each malfunction in sequence, recording cause and corrective action.
Materials: Full array of oxygen delivery devices (nasal cannula, simple mask, partial rebreather, non-rebreather, Venturi mask with color-coded adapters, high-flow nasal cannula system), oxygen flow meters, bubble humidifiers, manikin head, scenario cards, troubleshooting log sheet
Check: Instructor checklist for setup rotation: all five devices must be assembled with correct flow and fit; plus four-malfunction troubleshooting sheet graded — students must pass assembly on at least four of five devices and correctly diagnose all four malfunctions to pass.

Practice check8

Answer each question, then check it. Nothing is recorded; this is your own practice.

Clinical Application / Patient Care
1. A patient with known chronic CO₂ retention and an SpO₂ of 88% requires controlled oxygen therapy. Which device is most appropriate?
Equipment Troubleshooting / Patient Safety
2. During a routine check you find that a patient's non-rebreather bag collapses completely with each breath. What is the most immediate corrective action?
Equipment Safety / Patient Care
3. A simple mask is running at 4 L/min. Why is this a safety concern?
Mathematics / Equipment
4. Using the standard nasal cannula estimation formula, what is the approximate FiO₂ delivered at 5 L/min?
Clinical Decision Making / Patient Care
5. A patient scenario card states: SpO₂ 82% on room air, acute hypoxemic respiratory failure, cooperative, not at risk for CO₂ retention. Which device is the best initial choice?
Equipment Troubleshooting
6. While setting up a partial rebreather mask you notice the reservoir bag does not inflate. What is the most likely cause?
Equipment / Clinical Application
7. Which of the following is a correct statement about the Venturi mask?
Equipment Troubleshooting / Patient Care
8. A student assembles a bubble humidifier but notices no bubbling when oxygen is flowing. What is the most appropriate next action?