Objective. Students will insert a laryngeal mask airway (LMA) and at least one other supraglottic device on an airway trainer, confirming placement by ventilation compliance and auscultation.
This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.
VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.
Supraglottic airway devices (SGAs) sit above the glottis and create a seal around or near the laryngeal inlet. They are used when bag-valve-mask (BVM) ventilation alone is inadequate and when endotracheal intubation is not immediately possible or appropriate. In respiratory therapy practice, SGAs appear in emergency airway rescue, procedural sedation support, and intraoperative settings. Understanding their mechanics, correct sizing, insertion technique, and limitations is essential before using them on real patients.
Laryngeal Mask Airway (LMA) — The LMA is the most widely studied SGA. A flexible tube connects to an elliptical, inflatable cuff that seats against the perilaryngeal tissues, forming a low-pressure seal over the laryngeal opening. Classic LMAs are reusable; disposable versions are also common. The LMA does not protect against aspiration, because it does not pass through the vocal cords.
King LT (Laryngeal Tube) — The King LT is a single-use, dual-cuff device. It is inserted blindly into the esophagus; one cuff seals the esophagus and a second cuff seals the oropharynx. Ventilation occurs through side ports positioned at the laryngeal level. A single inflation port inflates both cuffs simultaneously.
i-gel — The i-gel uses a soft, non-inflatable thermoplastic cuff that conforms to perilaryngeal anatomy by body heat. No syringe is needed for cuff inflation. It also incorporates a gastric channel that allows passage of a suction catheter to decompress the stomach.
Choosing the correct size prevents seal failure and airway trauma. Manufacturer charts match device size to patient weight or age. For LMAs, common adult sizes are 3 (small adult), 4 (adult), and 5 (large adult). Pediatric sizes are based on weight in kilograms. Never estimate by eye alone — consult the sizing guide printed on the device packaging or the clinical reference posted in your skills lab.
1. Select and inspect the device. Choose the correct size; check the cuff for leaks by inflating and watching for symmetry, then fully deflate it before insertion.
2. Position the patient. Place the patient (or trainer) in the sniffing position — head extended, neck slightly flexed — unless a cervical spine injury is suspected.
3. Lubricate the posterior surface of the cuff with a water-soluble lubricant. Do not lubricate the anterior (grill) surface; lubricant entering the airway risks aspiration and laryngospasm.
4. Open the mouth. Use a scissor technique or jaw thrust. The device requires adequate mouth opening (typically ≥ 2 cm; check the manufacturer recommendation).
5. Insert the device. Hold the LMA like a pen at the tube-cuff junction. Press the deflated cuff tip against the hard palate and advance it in a smooth arc following the palatal curve, using your index finger as a guide, until firm resistance is felt.
6. Inflate the cuff. Use the correct volume of air as specified for the size (listed on the device label). Do not over-inflate; excessive pressure reduces seal quality and increases mucosal injury risk.
7. Connect the BVM and ventilate. Observe chest rise and deliver two test breaths.
8. Confirm placement. Auscultate the bilateral lung fields for equal breath sounds and over the epigastrium to rule out gastric ventilation. Observe end-tidal CO₂ waveform if capnography is available.
9. Secure the device. Tape or tie the tube to prevent displacement.
Placement confirmation relies on multiple methods used together:
If mouth opening is limited (e.g., trismus, clenched teeth), standard LMA insertion may be impossible because it requires a two-finger mouth opening. In that scenario, consider a smaller LMA size if any opening exists, or switch to a device with a smaller profile. The i-gel and King LT can sometimes be inserted with less mouth opening than a standard LMA. Document your clinical reasoning: note the observed limitation, the device you selected, and the outcome.
If a seal leak is evident after correct cuff inflation, reassess positioning and device size before adding more cuff air. Avoid exceeding the maximum cuff volume listed on the device.
Answer each question, then check it. Nothing is recorded; this is your own practice.